Link to ladder: L8UL201022
Module versions: M8UL2T0010220B2 | QR pdf
Workflow: | WF_2025_01_31_12_38_58.pdf |
Module: M8UL2T0010220B2 [ Site: GSI]
FEB_A: 1219 FEB_B: 2214-- Last status: 2025-03-07 15:28 [ Tested in Hk.Hl. ]
History of statuses:
Comments:
- 2025-03-18 15:33 : < Lady CollazoSanchez > ( 41 broken channels (39 on n-side & 2 on p-side) )
- 2025-03-18 15:28 : < Lady CollazoSanchez > ( Drying out and rearranging the microcables solved the breakdown. )
- 2025-02-13 14:17 : < Johann Heuser > ( presented by Anju on 22.01.2025 current increase from 100 V, does not reach 200 V )
- 2025-02-13 13:51 : < Johann Heuser > ( reported by Anju on 12.02.2025 passed with errors: M8UL2T0010220B2 Reason: 41 broken channels (39 on n-side and 2 on p-side) but current breakthrough at 370 V )
Change Module Status:
Ladder: L8UL201022
Module: M8UL2T0010220B2
Sensor: 10322
Status: Tested in Hk.Hl.