Link to ladder: L0DR100009
Module [ Site: GSI] versions:
M0DR1B4100094A2
| M0DR1B4000094A2 |
QR pdf
-- Last status: 2026-03-12 13:42 [ Dismissed ]
Desired grade: B
History of statuses:
Comments:
- 2026-03-12 13:42 { Johann Heuser } ( to be rebuilt )
- 2026-03-03 11:18 { Johann Heuser } ( during biasramp-up current fluctuations are seen from 280 V on . They remain visible in long-term biassing at 300 V and 250V. Fluctuations are by several microamperes on short time scales. )
- 2026-03-02 04:59 { Uli Frankenfeld } ( done: long-term IV test )
- 2026-02-25 15:27 { Johann Heuser } ( reconsidered: send module for long-term IV test. )
- 2026-02-25 15:18 { Johann Heuser } ( adjusted status - for module rebuilding )
- 2026-02-25 10:27 { Johann Heuser } ( Grade A sensor turned bad to D, in addition the IV curve has funny kink. Ask Oleksandr to resurrect sensor. Rebuild module. )
- 2026-01-20 11:40 { Anju Sharma } ( During IV it shows clear ohmic behavior after 170V. Current at EOL is 80 microA. )
Workflow
No documents uploaded yet.
⇧ Top
Change Module Status:
Ladder: L0DR100009
Module: M0DR1B4000094A2
Sensor: 17293