Link to ladder: L3DR200012
Module [ Site: GSI] versions: M3DR2T1000121B2 |
QR pdf
FEB_A: 1284 FEB_B: 2278 -- Last status: 2026-05-26 14:42 [ Undef. status at Hk.Hl. ]
Desired grade: A
History of statuses:
Comments:
- 2026-05-29 14:10 { Oleksandr Suddia } ( FEB "B" ASIC 7 - NEW )
- 2026-05-29 14:09 { Oleksandr Suddia } ( ASIC 7 - NEW )
- 2026-05-26 14:42 { Lady CollazoSanchez } ( N-side FEB: ASIC HW_6 no analog response. )
- 2026-05-26 14:40 { Lady CollazoSanchez } ( P-side FEB: issues with one od the two 1.2 V LDO affecting 4 ASICs. Output 1.2 V LDO for ASICs w/ HW addresses 0,2,4,6 is very low. AFE of these group of ASICs are not properly powered, CSA are off. No analog response. Overall 1.2 V line consumption is 1.3 A )
- 2026-05-26 14:24 { Lady CollazoSanchez } ( During IV, the module shows an slow increased leakage current after 360 V, reaching the 59uA at 500 V. )
- 2026-04-28 10:16 { Kerstin Schuenemann } ( ODI all ok )
No documents uploaded yet.
⇧ Top
Change Module Status:
Ladder: L3DR200012
Module: M3DR2T1000121B2
Sensor: 07102
Status: Undef. status at Hk.Hl.