Link to ladder: L3UR401113
Module [ Site: KIT] versions: M3UR4B0011130B2 |
QR pdf
-- Last status: 2026-03-31 09:31 [ Tested in Hk.Hl. ]
Desired grade: C
History of statuses:
Comments:
- 2026-04-01 15:03 { Gnana Sindhu Subramanya } ( N-side: ASIC 0_UL0_P, and ASIC 0_UL0_N have globtop And ASIC 1_UL1_N shows 50 Ω ASIC 7_UL0_P, ASIC 7_UL0_N, ASIC 7_UL1_P, ASIC 7_UL1_N have glob top P-side : ASIC_HW 0, Uplink is missing during the functional test, but ODI is okay for this ASIC HW 0 Module needs 1 uplink. ENC looks good. )
Workflow not found !!!
No documents uploaded yet.
⇧ Top
Change Module Status:
Ladder: L3UR401113
Module: M3UR4B0011130B2
Sensor: 01142
Status: Tested in Hk.Hl.