Desired grade: A
History of statuses:
Comments:
- 2026-03-26 11:41 { Gnana Sindhu Subramanya } ( ODI is okay )
- 2026-03-16 16:38 { Johann Heuser } ( ASIC was replaced. Full test in HL needed. )
- 2026-03-12 13:22 { Oleksandr Suddia } ( Replaced ASIC1_HW6 )
- 2026-03-05 13:02 { Uli Frankenfeld } ( ASIC exchange scheduled )
- 2026-02-25 16:08 { Johann Heuser } ( Oleksandr to replace one chip )
- 2026-02-04 11:34 { Anju Sharma } ( Sorry for previous comments. Corrected: Low ODI (1.8kohms, both uplinks, HW_ID 6) )
- 2026-02-04 11:22 { Anju Sharma } ( ODI is okay )
- 2026-02-02 16:31 { Anju Sharma } ( Retested )
- 2025-03-18 15:26 { Lady CollazoSanchez } ( Dry out and rearranging of the microcables solved the breakdown )
- 2025-02-13 14:25 { Johann Heuser } ( B but required A )
- 2025-02-13 13:37 { Johann Heuser } ( presented by Anju on 12.02.2025 --> breakdown at > 350 V --> maybe class B (from A) )
- 2025-02-11 11:30 { Anju Sharma } ( Sharp breakdown at 370V )
- 2025-02-11 11:29 { Anju Sharma } ( Sharp breakdown at 370V )
- 2025-02-11 11:28 { Anju Sharma } ( Sharp breakdown at 370V )
- 2025-02-11 11:27 { Anju Sharma } ( Sharp breakdown at 375V (Optical grade A) )
No documents uploaded yet.
⇧ Top
Change Module Status:
Ladder: L8UL201022
Module: M8UL2T2010222B2
Sensor: 18054
Status: Tested Uplinks